How to Measure Intraocular Pressure
Intraocular pressure concerns the fluid within the eye that exerts pressure on the inner wall of the eyeball. It is normally between 10 and 20 millimeters of mercury. An intraocular pressure above this range can cause glaucoma. An applanation tonometer is the standard method of measuring intraocular pressure.
Items Needed
- Eye examination chair
- Slit lamp microscope
- Tonometer
- Anesthetic eye drops
- Fluorescein eye drops
- Blue filter
- Adjust the height of the examination chair so that the patient can comfortably place his chin into the chin cup. The patient’s head must also be against the chair’s headband.
- Move the tonometer’s prism so that the “0” mark on the prism aligns with the mark on the tonometer’s head. Turn the calibration dial on the tonometer to 10 mmHg.
- Set the magnification factor on the slit microscope to 10. Use anesthetic eye drops to deaden the sensation in the patient’s eyes. Administer fluorescein eye drops to make surface of the eyeball more visible.
- Place a blue filter on the slit lamp microscope so you can see the fluorescein. Hold the tonometer to the side of the patient’s head and shine the lamp of the slit lamp microscope into the head of the tonometer so that it reflects into the patient’s eyes. Instruct the patient to look straight ahead, and lift the patient’s eyelid so that you don’t put pressure on the patient’s eye.
- Hold the tonometer so that it is perpendicular to the eye, and move it forward until it touches the eye. Turn the calibration bow of the tonometer clockwise until the two semicircles visible through the tonometer’s head touch to form the letter “S."
- Read the calibration dial on the tonometer to obtain the patient's intraocular pressure.